Spectrometers & sensors
SE-REDTIDE-UV-CUVETTE
Ready to go UV-VIS absorbance spectrophotometer system for $2,890

The RedTide-UV-Cuvetteis a low-cost, small- footprint lab spectrophotometer that's ideal as a general-purpose instrument for budget-conscious teaching and research labs. The system consists of the following”
Red Tide Spectrometer. This versatile low cost unit has has a wavelength range of 200-850 nm, and utilizes a detector with 650 active pixels; that’s 650 data points in one full spectrum, or one data point per nanometer. Configured with a 25 µm entrance slit, the Red Tide offers ~2.0 nm optical resolution (FWHM). The Red Tide plugs into you computer through a USB port.
The spectrometer is coupled to our USB-ISS-UV-VIS combination sample holder and deuterium tungsten light source (200-1100 nm). This sampling system holds standard 1-cm cuvettes for easy-to-perform absorbance measurements. The USB-ISS-UV-VIS allows you to control both the intensity of the tungsten bulb and a shutter via software. The USB-ISS-UV-VIS requires an external power supply (included).
Spectrasuite Software works on Mac, Linux or Windows based computers. Spectrasuite is loaded with features: real time absorbance spectra, graphical functions for annotations and printing, time series control and strip chart mode, peak finder, automatic wavelength calibration, mapping to absorbance standards, and much more.
CVDUV1S You get one pack of 100 disposable UV cuvettes. These cuvettes have 4 clear sides, are 1.5 – 3.0 ml volume and a1cm pathlength,
Specifications:
Physical | |
Spectrometer dimensions: |
89.1 mm x 63.3 mm x 34.4 mm |
Spectrometer weight: |
190 grams |
Light source dimensions: |
198 x 105.1 x 40.6 |
Light source weight: |
200 g |
Detector Specifications | |
Detector: |
Sony ILX-511 Linear CCD Detector |
Pixels: |
650 pixels |
Pixel size: |
14 µm x 200 µm |
Pixel well depth: |
62,500 electrons |
Signal-to-noise ratio: |
250:1 (at full signal) |
A/D resolution: |
12 bit |
Corrected linearity: |
>99.8% |
Sensitivity: |
75 counts at 400 NM; 41 photons/count at 600 nm |
Optical Bench | |
Design: |
f/4, Asymmetrical crossed Czerny-Turner |
Focal length: |
42 mm input; 68 mm output |
Entrance aperture: |
25 µm wide slit |
Grating: |
Grating #1 |
OFLV filter: |
OFLV-200-850 |
UV enhanced window: |
Yes, UV4 |
Fiber optic connector: |
SMA 905 to 0.22 numerical aperture single-strand optical fiber |
Spectroscopic | |
Wavelength range: |
200 to 880 nm |
Optical resolution: |
2.0 nm FWHM |
Signal-to-noise ratio: |
300:1 (at full signal) |
A/D resolution: |
12 bit |
Dark noise: |
3.2 RMS |
Integration time: |
3ms to 65 seconds |
Dynamic range: |
2 x 108 (system), 1300:1 for a single acquisition |
Stray light: |
<0.05% at 600 nm; 0.10% at 435 nm |
Light Source/Sample Holder | |
Power consumption: |
1.8 A @ 5 VDC; external power supply (included) |
Pathlength: |
1 cm |
Cuvette shape: |
Square |
Light source: |
Deuterium tungsten |
"Z" dimension: |
15 mm |
Bulb life (hours): |
800 deuterium; 2,000 tungsten |
Time to stabilize: |
~30 minutes |
Electronics | |
Power consumption: |
250 mA @ 5 VDC |
Data transfer speed: |
Full scans to memory every 4 ms with USB 2.0 port, 18 ms with USB 1.1 port |
Inputs/Outputs: |
Yes, 8 onboard digital user-programmable GPIOs |
Analog channels: |
No |
Auto nulling: |
Yes |
Breakout box compatible: |
Yes, HR4-BREAKOUT |
Trigger modes: |
5 modes |
Strobe functions: |
Yes |
Connector: |
22-pin connector |
Computer | |
Operating systems: |
Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
Computer interfaces: |
USB 2.0 @ 480 Mbps (USB1.1 compatible); RS-232 (2-wire) @ 115.2 K baud |
Peripheral interfaces: |
SPI (3-wire); I2C inter-integrated circuit |