Spectrometers & sensors
USB2000+ fast gated CCD array spectrometer
Â

USB2000+
 Miniature Fiber Optic Spectrometer
The USB2000+ Miniature Fiber Optic Spectrometer is all about speed -- a powerful 2-MHz analog-to-digital (A/D) converter, programmable electronics, a 2048-element CCD-array detector that is gateable, and a high-speed USB 2.0 port. The USB2000+ captures a full spectrum into memory every millisecond when the spectrometer is interfaced to a computer via a USB 2.0 port. This fast, gateable array is ideal for use in pulsed laser induced emission applications like LIBS and phosphorersence, as well as routine use for absorbance, reflection, or flourescence measurements.
1,000 Full Spectra/Second
The USB2000+ has an onboard programmable microcontroller that provides flexibility in controlling the spectrometer and accessories. Through a 20-pin connector, you can implement all operating parameters in the software, such as controlling external light sources, creating processes and routines and retrieving data from external devices. The USB2000+ gives you access to 10 user-programmable digital I/Os for interfacing to other equipment; one analog input and one analog output; and a pulse generator for triggering other devices. (Programming the I/Os requires SpectraSuite Spectroscopy Operating Software.)
Setting up the USB2000+ Spectrometer is easy. The user simply installs our Spectrometer Operating Software and then connects the USB cable from the spectrometer to the computer. Wavelength calibration coefficients unique to each spectrometer are programmed into a memory chip right on the USB2000+; the software simply reads these values from the spectrometer. What's more, the USB2000+ requires no external power supply; it draws its power from the computer. In fact, the USB port can be used to power light sources that connect to the spectrometer.Â
The USB2000+ couples a low-cost, high-performance 2048-element linear CCD-array detector with an optical bench that's small enough to fit into the palm of your hand -- the same detector and optical bench that have provided superior performance to the users of our 90,000+ systems out in the field. The USB2000+ also works the same way as other Ocean Optics spectrometers in that it accepts light energy transmitted through single-strand optical fiber and disperses it via a fixed grating across the linear CCD array detector, which is responsive from 200-1100 nm.
Just as with all other Ocean Optics user-configured spectrometers, users work with our applications scientists when selecting the USB2000+ 14 gratings, 6 slits and hundreds of fiber optic accessories such as light sources, probes and optical fibers to create the optimal system for their application.
Â
Specifications
Physical | |
Dimensions: |
89.1 mm x 63.3 mm x 34.4 mm |
Weight: |
190 grams |
Detector | |
Detector: |
Sony ILX511 linear silicon CCD array |
Detector range: |
200-1100 nm |
Pixels: |
2048 pixels |
Pixel size: |
14 µm x 200 µm |
Pixel well depth: |
~62,500 electrons |
Sensitivity: |
75 photons/count at 400 nm; 41 photons/count at 600 nm |
Optical Bench | |
Design: |
f/4, Symmetrical crossed Czerny-Turner |
Focal length: |
42 mm input; 68 mm output |
Entrance aperture: |
5, 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit) |
Grating options: |
14 different gratings, UV through Shortwave NIR |
HC-1 grating option: |
No |
Detector collection lens option: |
Yes, L2 |
OFLV filter options: |
OFLV-200-850; OFLV-350-1000 |
Other bench filter options: |
Longpass OF-1 filters |
Collimating and focusing mirrors: |
Standard or SAG+ |
UV enhanced window: |
Yes, UV2 |
Fiber optic connector: |
SMA 905 to 0.22 numerical aperture single-strand optical fiber |
Spectroscopic | |
Wavelength range: |
Grating dependent |
Optical resolution: |
~0.3-10.0 nm FWHM |
Signal-to-noise ratio: |
250:1 (at full signal) |
A/D resolution: |
16 bit |
Dark noise: |
3.2 RMS counts |
Dynamic range: |
2 x 10^8 (system); 1300:1 for a single acquisition |
Integration time: |
1 ms to 65 seconds |
Stray light: |
<0.05% at 600 nm; <0.10% at 435 nm |
Corrected linearity: |
>99.8% |
Electronics | |
Power consumption: |
90 mA @ 5 VDC |
Data transfer speed: |
Full scans to memory every 1 ms with USB 2.0 or 1.1 port, 300 ms with serial port |
Inputs/Outputs: |
No |
Analog channels: |
No |
Auto nulling: |
No |
Breakout box compatibility: |
No |
Trigger modes: |
3 modes |
Strobe functions: |
Yes |
Gated delay feature: |
Yes |
Connector: |
10-pin connector |
Computer | |
Operating systems: |
Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
Computer interfaces: |
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 57.6 K baud |
Peripheral interfaces: |
I2C inter-integrated circuit |